The agreement study is underway

Illustration: a close view of one narrowed coronary vessel on an angiogram with two outlines, the expert annotation as a white dashed line and the application's contour in yellow, tracking each other closely.
Illustration: ARCADE public dataset, expert annotation and APEC Vox QCA contour

We are measuring our output against the commercial 3D-QCA analyses in the validation cohort — minimum lumen diameter, reference diameter, %diameter stenosis and lesion length, case by case, reported as bias and limits of agreement.

This is the gate. Nothing downstream means anything until it passes, the physiology work included, and %stenosis stays a research output until it does.

The first findings are already negative in places: the pipeline under-reads where the lumen pinches to one or two pixels. That is what the study is for. When it reports, the result goes to the top of the publications page whichever way it comes out.